Category Calibrators (RTD)
Resistance Range 10 Ω to 3200Ω
Weight 587g
Resistance Resolution 0.1Ω
Height 52mm
Length 98mm
Overview
Fluke 712 RTD Calibrator
A handheld single function unit in the familiar Fluke 80 Series DMM package. These calibrators are compact and lightweight with easy-to-use pushbutton interface.
Clear 5 digit LCD display
Measure temperature from RTD output
Simulate RTD output
Operate with 7 types of RTD: 100 200 500 1000;JIS and Minco 3916; Ni 120
Measure additional RTDs using Ohms function
Simulate additional RTDs using Ohms source function
°F or °C selectable
4 shrouded banana jacks
Operating temperature -10 to +55°C
Technical specification
Temperature range +18 to +28°C
Function Range Resolution Accuracy
Resistance 10Ω to 3200Ω 0.1Ω ±0.1Ω
RTD measure 200 to 800°C
(Pt 100) 0.1°C, 0.1°F ±0.1Ω
RTD simulate 200 to 800°C
(Pt 100) 0.1°C, 0.1°F ±0.1Ω
(0.5 to 2mA)
±0.15Ω
Dimensions (L x W x D) 201 x 98 x 52mm
Weight 600g (0.15 to 0.5mA)
Supplied with
protective yellow holster, test leads and crocodile clips, 9V alkaline battery (installed) and an instruction leaflet.
|